Range
Current Limit
100 to 500 VDC
20mA
500 to 4000 VDC
35mA
4000 to 4500 VDC
30mA
4500 to 5000 VDC
25mA
5,000 to 7,000 VDC*
5mA
* Requires option DC-7
DC Current Limits with option DC-10:
Range
Current Limit
100 to 10,000 VDC
10mA
HIGH RESISTANCE MODE
(Insulation Resistance)
Resistance Range
Minimum Resistance
Test Voltage
2.6G
Ω
100K
Ω
100VDC
13G
Ω
500K
Ω
500VDC
26G
Ω
1M
Ω
1,000VDC
65G
Ω
2.5M
Ω
2,500VDC
Accuracy:
Combined uncertainty of DC Voltage Current Measurement Accuracy
LOW RESISTANCE MODE
(Earth Continuity)
Full Scale
8.5
Ω
Maximum including lead resistance.
Resolution
1m
Ω
Test Current
50mA
Lead Compensation
Automatic offset circuit "nulls" out two wire lead resistance
Accuracy
0.3% of r 30m
Ω
(up to 5
Ω
)
General Specifications
Arc Detection Sensitivity
User selectable 3 microseconds to 30 microseconds
Display
40 character alphanumeric LCD
High Voltage Indicator
Lightning bolt illuminated by independent output voltage monitor.