Audible alarm sounds prior to voltage output.
Voltage Resolution
1 volt AC or DC
Dwell Time Accuracy
+2 seconds/minute, -500mS/minute
Slew Rate
10 volts/second to 1999 volts/second, user selectable
Limits
Test automatically aborts when digitally stored minimum or
maximum limit is exceeded.
Remote Interface Option "IF-4"
GPIB - IEE-488.2 Subset compatible
RS232C - 1200, 2400, 4800, 9600 Baud rate selectable
Parallel - IBM/Centronics compatible
VICL - Vitrek auxiliary unit control loop
Accuracy
Stated specifications are bipolar and apply for one year within ± 5°C
from calibration temperature (20 to 25°C).
Temperature Coefficient
0.1 times stated accuracy per °C beyond calibration temperature
window
Calibration
Traceable to the National Institute of Standards and Technology
(NIST). Automatic "Covers on" calibration, internal or external.
Calibration constants stored in non-volatile memory.
Warm up Time
No warm up time is required for units stored in ambient
temperatures ranging from 64°F to 82°F (18°C to 28°C). A
minimum 30 minute warm up time is recommended for units stored
in temperatures outside that range. A minimum 60 minute warm up
time is recommended prior to performing an external calibration
procedure.
Operating Temperature
0°C to 50°C
Humidity
90% RH, 0° to 40°C
Power
115/230 VAC ± 10%, 350 VA Max
Dimensions
89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D)
Weight
14Kg (34 lb.) Net / 18Kg (40 lb.) Shipping