The display now shows the main configuration menu, an example of which is
–
TEST…
PRINTOUT…
SYSTEM…
INTERFACES…
DIGITAL I/O…
BUILD…
SET TO DEFAULTS…
LOCK PASSWORD:000000
RELOCK…
•
Using the Left Arrow or Right Arrow keys as needed, change the selection point to the INTERFACES line.
•
Press the
ENTER
key.
•
The 95x now displays the base interfaces configuration settings menu, an example of which is
–
RS232 BAUD: 115200
SWITCH: NONE
DISABLE USB: NO
DISABLE GPIB: NO
DISABLE ENET: NO
ETHERNET…
GPIB ADDR: 2
•
The user may navigate this sub-menu and change the displayed settings as required. The settings
affecting using the RS232 to control the 95x are
–
o
DISABLE ENET
. This must be set to NO to enable the Ethernet interface in the 95x.
o
ETHERNET
… This selection is the entry point to the sub
-menu for configuring the Ethernet
interface. The user should navigate to select this menu line and press then
ENTER
key.
•
The Ethernet sub-menu allows the user to configure the 95x for the local network. An example of this
sub-menu is as follows
–
•
USE DHCP: YES
IP: 000.000.000.000
SNET:000.000.000.000
GTWY:000.000.000.000
MAC: 1234567890AB
•
The user may navigate this sub-menu and change the displayed settings as required. The settings are
–
o
USE DHCP
. Allows the user to configure the 95x to use DHCP for IP address allocation (YES) or
not (NO).
o
IP
. If using DHCP, this shows the IP address which the 95x has been allocated. If not using DHCP
this allows the user to enter an IP address for the 95x on the local network.
o
SNET
. If using DHCP, this shows the IP subnet mask which the 95x has been given. If not using
DHCP this allows the user to enter the subnet mask for the local network.
o
GTWY
. If using DHCP, this shows the IP of the gateway to additional networks which the 95x has
been given. If not using DHCP this allows the user to enter the IP address of the gateway.
o
MAC
. This is not editable and shows the MAC address of the 95x. This may be needed to
configure the local network to allow connection of the 95x to it. In most networks this is not
needed.