95x Series Operating Manual - May 17, 2022
Page
12
of
155
computer is required. The 95x is capable of controlling switch matrix units (also available from Vitrek)
–
up to 256 channels can be controlled without needing a computer or software. The 95x can even print a
test report on a printer for you when a test sequence has completed.
•
System operation - Wide range of interfaces available. If the user wishes to use the 95x with a computer,
then RS232, GPIB or Ethernet interfacing can be chosen as the interfacing medium between them. Giving
the user the flexibility to use the 95x in almost any computing environment. Software (QuickTest Pro) is
available from Vitrek to provide all the control needed for any system from the simple (just the 95x) to the
most complex with the 95x terminals being multiplexed between DUTs and/or points within DUTs by up
to 1024 channels in switch matrix units (also available from Vitrek).
•
High Speed. The 95x is capable of performing very quickly, up to 100 tests per second can be performed
with the test results being made available both during each test and after the entire sequence has been
run. The 95x is also capable of chaining similar tests without needing to reduce the applied voltage or
current to zero between test steps, this considerably speeds up testing when multiple test levels are
required.
•
All of the measurements, all of the time. The 95x does not just measure what the user has set limits for,
all measurement results for the specific type of test being performed are made available to the user.
•
Energy Efficient. The 95x uses direct line switching power supplies to provide a very energy efficient
instrument, the 95x only draws significant power from the line when needed to power the load.
AVAILABLE MODELS AND OPTIONS
AC Voltage Testing
DC Voltage Testing AC Low Resistance Testing
951i
20-6000V
20-6500V
No
951i
952i
0.1-40A @ 8V
952i
953i
40-11000V
No
953i
954i
0.1-40A @ 8V
954i
955i
40-10000V
No
955i
956i
No
20-6500V
No
956i
957i
20-6000V
(40-10000V Opt AC10)
75-15000V
No
(0.1-40A Opt GB40)
957i
959i
No
No
0.1-40A @ 8V
959i
In addition to the above, all models have 2- and 4-wire DC Low Resistance and Ground Leakage Testing capabilities,
a RS232 interface, a Digital I/O interface and a VICL proprietary interface.
INTERFACING OPTION
This option may be fitted in any 95x.
Option
GPIB-9
adds a GPIB interface.
The 95x has Digital I/O, RS232, Ethernet, and USB Printer interfaces as standard.
DC VOLTAGE OPTIONS
This option may be fitted in any 95x except the 959i.
Option
DCNEG
changes the polarity of the DC Voltage capability to negative but limits the minimum DC output
voltage to 200V.
INTERNAL AC VOLTAGE OPTIONS
One or none of these options may be fitted in a 951i, 952i, 953i, 954i or 957i.