INPHSA - 1
st
checked value is In-Phase leakage current
QUADA - 1
st
checked value is Quadrature leakage current
RMSO - 1
st
checked value is RMS leakage impedance
INPHSO - 1
st
checked value is In-Phase leakage impedance
QUADO - 1
st
checked value is Quadrature leakage impedance
8
<NR3>
1
st
Minimum Leakage Limit (in A or
Ω
)
<EMPTY>
Only valid if no 1
st
checked leakage
9
<NR3>
1
st
Maximum Leakage Limit (in A or
Ω
)
<EMPTY>
No 1
st
maximum limit (only valid for impedance or no 1
st
checked leakage)
10
<STRING>
NONE - No 2
nd
checked leakage
RMSA - 2
nd
checked value is RMS leakage current
INPHSA - 2
nd
checked value is In-Phase leakage current
QUADA - 2
nd
checked value is Quadrature leakage current
RMSO - 2
nd
checked value is RMS leakage impedance
INPHSO - 2
nd
checked value is In-Phase leakage impedance
QUADO - 2
nd
checked value is Quadrature leakage impedance
11
<NR3>
2
nd
Minimum Leakage Limit (in A or
Ω
)
<EMPTY>
Only valid if no 2
nd
checked leakage
12
<NR3>
2
nd
Maximum Leakage Limit (in A or
Ω
)
<EMPTY>
No 2
nd
maximum limit (only valid for impedance or no 2
nd
checked leakage)
13
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
14
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 13
15
<STRING>
NONE - No discharge if next step compatible
FAST - Fast discharge
RAMP - Discharge same as ramp
16
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
17
<EMPTY>
Option AC-30 is not used, may only be <MISSING> if field 18 is also <MISSING>
<MISSING>
<STRING>
INT : Option AC-30 is not used
EXT : Option AC-30 is to be used (valid if option AC-30 is fitted)
18
<EMPTY>
Load is not grounded
<MISSING>
<STRING>
GND : Load is grounded (valid if option HSS is fitted)
ISO : Load is not grounded
Example
–
ADD,ACW,1000.0,60.0,0.01,1.5,60.0,RMSA,0.0,0.005,NONE,,,,,FAST,ABORT
Configures the following (in order)
–
An ACW type test step
1000V test voltage
60Hz test frequency
10mApk breakdown limit
1.5 second ramp time
60 second dwell time