95x Series Operating Manual - May 17, 2022
Page
22
of
155
b.
PASS
. This is illuminated whenever a test sequence is passing (while running a test sequence), or
it has passed all test steps (following completion of the test sequence).
c.
FAIL
. This is illuminated whenever a test sequence is failing (while running a test sequence) or
the previously run test sequence has failed any test steps (following completion of the test
sequence).
d.
TESTING
. This is illuminated when the 95x is running a test sequence.
e.
READY
. This is illuminated after the user has selected a test sequence to run and the 95x is ready
to perform that test sequence.
f.
ARC
. This is illuminated whenever the 95x is detecting an arc current while running a test
sequence and the specific test step is enabled to detect arc current.
g.
C/V LIMITED
. This is illuminated when the 95x output is either current or voltage limited by the
load while running a test sequence. This is only used in certain types of test steps.
h.
REMOTE
. This is illuminated when the 95x is under the control of an interface (i.e., RS232,
Ethernet or GPIB). If the user wishes to return to front panel control of the 95x the CNFG key
should be pressed to achieve this.
7.
The menu keys. These initiate a menu allowing the user to perform certain activities via the front panel of
the 95x. Many of these can be disabled by the user by requiring a password in order to utilize them.
a.
SELECT
. Initiates a menu allowing the user to select a test sequence which has already been
defined in the 95x. This never requires a password.
b.
EDIT
. Initiates a menu allowing the user to select an existing test sequence and then edit it.
After editing it is automatically made ready to run. This optionally requires a password.
c.
NEW
. Initiates a menu allowing the user to select a presently undefined test sequence number
and name, and then to create the sequence of tests to be performed. This optionally requires a
password. While editing or creating a test sequence, this key is also used to insert a new test
step into the test sequence.
d.
DEL
. Initiates a menu allowing the user to select an existing test sequence and delete it from the
95x. This optionally requires a password. While editing or creating a test sequence, this key is
also used to delete a test step from the test sequence.
e.
. If option UL-2 or GUL-3 is fitted and a suitable printer is attached, this key causes a
printout to be generated. This never requires a password. The printout generated is dependent
on the menu (if any) which is active when the key is pressed
–
i.
If pressed while no menu is active, or the configuration menu is active, a configuration
printout is generated.
ii.
If pressed while selecting an existing test sequence, a test sequence printout is
generated.
iii.
If pressed while reviewing the results of a completed test sequence, a test results report
printout is generated. The 95x can also be configured to automatically produce this
printout if desired.
f.
VIEW
. Initiates a menu allowing the user to either
–