95x Series Operating Manual - May 17, 2022
Page
33
of
155
or never allow it (DISABLED). If fast rerun is not enabled, the user must terminate reviewing
results by using the
STOP
button before they can rerun the test sequence with the
START
button.
o
USER MAX DCV
. This allows the user to set a maximum DC Voltage which the 95x will produce.
This limits all test steps subsequently defined and will not allow any existing test sequence
containing a test step level outside of this limit to run. Note, the factory default setting for this is
10KV, some models and options may allow a higher setting than this.
o
USER MAX ACV
. This allows the user to set a maximum AC Voltage which the 95x will produce.
This limits all test steps subsequently defined and will not allow any existing test sequence
containing a test step level outside of this limit to run. Note, the factory default setting for this is
10KV, some models and options may allow a higher setting than this.
o
HI SAFETY
. This turns on/off the ability of the 95x to detect excessive
HV
terminal current in high
voltage test step types (see the specific test type sections of this manual for details).
o
MIN LOAD
. This turns on/off the ability to detect that there is a minimum load attached during
DCW or DCIR type test steps (see the specific test type sections of this manual for details). If
enabled in this menu, then each DCW and DCIR test step may be individually configured for
minimum loading detection as needed.
o
MAX DISCHARGE
. This sets the maximum current which will be used by the 95x when
discharging a load at the end of a DCez, DCW or DCIR type test step. A value between 1 and
200mA may be configured.
•
When finished, press the
EXIT
key to exit this sub-menu, and then press
EXIT
again (if no additional
configuring is to be performed) to exit the main configuration menu and save the settings.
CREATING A NEW TEST SEQUENCE
•
Ensure that the display indicates that the 95x is in the base menu state, i.e., the display shows the date
and time. If necessary, press the
STOP
button to abort a menu and return to the base menu state.
•
Press the
NEW
key.
•
The display now shows the lowest presently empty test sequence number and a blank name field. An
example is -
NEW TEST SEQUENCE
# 2
•
Either change the test sequence number (using the numeric keys or the Up Arrow and Down Arrow keys)
or press the
ENTER
key. The display now selects the name field. The user can enter a name by pressing
the Up Arrow key followed using the Up, Down, Left and Right Arrow keys to change/select each
character, and the user may use the numeric keys to directly enter numeric characters.
•
Press the
ENTER
key. The display now shows the first test step (step #1) with no test type selected (NEW
is displayed). An example is -
Seq 2 Step 1 NEW
•
Using the Up or Down Arrow keys, select the desired first test step type.
•
Using the information for the selected test step type shown later in this document, program the
requirements for this test step.
•
If no more test steps are required, press the
EXIT
key. The 95x stores the test sequence and makes it
ready to be run.