95x Series Operating Manual - May 17, 2022
Page
94
of
155
Seq 1 Step 1 PAUSE
TIME: 0.00sec
•
TIME
. Allows the user to set the desired pause time. Any value between 0.00 and 9999 seconds may be
entered.
An example HOLD type menu is as follows
–
Seq 1 Step 1 HOLD
TIMEOUT: 30.0sec
•
TIMEOUT
. Allows the user to set the desired hold timeout. Any value between 0.00 and 9999 seconds
may be entered. Setting a zero value disables the timeout entirely; the 95x will wait “forever” for user
interaction to occur.
EXAMPLES
Example 1 -
The user wishes to program a 1 second pause into a test sequence to allow for wiring settling between two steps.
A PAUSE step is configured as
–
Seq 1 Step 28 PAUSE
TIME: 1.00sec
Example 2 -
The user wishes to program a step to have the sequence wait for user interaction before proceeding. This will be
used to allow the user to manually alter the DUT wiring. It is desired to timeout the sequence if the user does not
respond within 1 minute.
A HOLD step is configured as
–
Seq 1 Step 28 HOLD
TIME: 60.0sec
SPECIFICATIONS
Delay or Timeout
0.00 to 9999sec
<(± 0.05% ± 20ms) accuracy