95x Series Operating Manual - May 17, 2022
Page
139
of
155
Minimum load during ramp is set to 0 since both fields 15 and 16 are missing
DUT is isolated since both fields 15 and 16 are missing
GBEZ CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
GBEZ
2
<NR3>
Test Current (in Arms)
3
<NR3>
Test Frequency (in Hz)
4
<NR3>
Dwell time (in seconds)
<EMPTY>
Step is to be user terminated
5
<NR3>
Minimum Voltage Drop Limit (in Vrms)
6
<NR3>
Maximum Voltage Drop Limit (in Vrms)
7
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example
–
ADD,GBEZ,25.0,60.0,60.0,0.0,0.1,ABORT
Configures the following (in order)
–
A GBez type test step
25A test current
60Hz test frequency
60 second dwell time
No minimum voltage drop
0.1V maximum voltage drop
Abort test sequence if fails
GB CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
GB
2
<NR3>
Test Current (in Arms)
3
<NR3>
Test Frequency (in Hz)
4
<NR3>
Maximum open circuit drive (in Vrms)
5
<NR3>
Ramp Time (in seconds)
6
<NR3>
Dwell time (in seconds)
<EMPTY>
Step is to be user terminated
7
<STRING>
RMSV - Checked value is RMS voltage drop
INPHSV - Checked value is In-phase voltage drop
QUADV - Checked value is quadrature voltage drop
RMSO - Checked value is RMS impedance
INPHSO - Checked value is in-phase impedance
QUADO - Checked value is quadrature impedance
8
<NR3>
Minimum Limit (in V or
Ω
)
9
<NR3>
Maximum Limit (in V or
Ω
)
10
<STRING>
NONE - No discharge if next step compatible
FAST - Fast discharge