TEST STEP STATUS FLAGS
This is a <NR1> value formed by the addition (or logical OR) of the values shown in the table below. Note that a
zero total value indicates that no failure occurred.
Bit
#
Value
Description
0
1 95x Internal Fault
1
2 Unstable load
2
4 Breakdown detected (> setting)
3
8 HOLD Step timeout exceeded
4
16 User abort
5
32 Continuity Sense failure detected
6
64 Wiring fault detected
7
128 Arc detected
8
256 1
st
check <minimum setting
9
512 1
st
check >maximum setting
10
1024 2
nd
check <minimum setting
11
2048 2
nd
check >maximum setting
12
4096 INTERLOCK failure
13
8192 HV TRIP activated
14
16384 Switch Matrix unit did not communicate
15
32768 Breakdown detected (> surge drive ability limit for test type)
16
65536 95x Overheated
17
131072 Step type incompatible with the model and option content
18
262144 Minimum Loading was not encountered during ramp in a DCW or DCIR type step
19
524288 Breakdown detected (> sustained drive ability limit for test type)
20
1048576
A steady or decreasing current was not detected during dwell but the load was otherwise within
the limits in a DCIR type step configured to terminate in this manner
21
2097152 User forced abort