TEST SEQUENCE CONFIGURATION
There are settings in the 95x which apply to all test steps in all test sequences. To view or change these settings
perform the following actions
–
•
Ensure that the display indicates that the 95x is in the base menu state, i.e., the display shows the date
and time. If necessary, press the
STOP
button to abort a menu and return to the base menu state.
•
Press the
CNFG
key.
•
The display now shows the main configuration menu, an example of which is
–
TEST…
PRINTOUT…
SYSTEM…
INTERFACES…
DIGITAL I/O…
BUILD…
SET TO DEFAULTS…
LOCK PASSWORD:000000
RELOCK…
•
Using the Left Arrow or Right Arrow keys as needed, change the selection point to the TEST line.
•
Press the
ENTER
key.
•
The display now shows the test configuration menu. An example of the test configuration menu is as
follows
–
START BEEP: MED
PASS BEEP: MED
FAIL BEEPS: LOUD
ARC: FAIL ON DETECT
CONTINUITY SENSE:OFF
FAST RERUN: ENABLED
USER MAX DCV: 10.0KV
USER MAX ACV: 10.0KV
HI SAFETY: ENABLED
MIN LOAD: ENABLED
MAX DISCHARGE: 200mA
•
The user may navigate this sub-menu and change the displayed settings as required.
o
START BEEP
. This allows the user to set the volume of the beep sound emitted by the 95x when
a test sequence is started.
o
PASS BEEP
. This allows the user to set the volume of the beep sound emitted by the 95x when a
test sequence is completed with a PASS status.
o
FAIL BEEP
. This allows the user to set the volume of the beep sounds emitted by the 95x when a
test sequence is completed with a FAIL status.
o
ARC
. This allows the user to select whether all test steps configured for arc detection will either
fail the test if an arc is detected (FAIL ON DETECT) or not (DETECT ONLY). Note that arc detection
is enabled or disabled by a separate user setting in each individual test step.
o
CONTINUITY SENSE
. This allows the user to turn on/off the Continuity Sense feature in certain
types of test steps (see the specific test step type sections of this manual for details).
o
FAST RERUN
. This allows the 95x to always allow the user to rerun a test sequence without
having to terminate reviewing the results (ENABLED), or only allow it if all tests passed (IF PASS),