95x Series Operating Manual - May 17, 2022
Page
37
of
155
In-Phase Resistance (
Ω
inp)
Quadrature current (Aqua)
Quadrature Voltage (Vqua)
Quadrature impedance (
Ω
qua)
Capacitance (F)
Dissipation Factor (DF)
Peak or breakdown current (Apk)
Arc current (Aarc)
•
Many results are capable of having a negative polarity; a negative result is preceded by the
–
character. A
positive result has no preceding polarity character.
•
The 95x automatically formats the displayed result to give the best meaningful resolution using a
multiplier character as needed (e.g., G, M, K, m, u, n, or p). Displayed results may have slightly limited
resolution; full resolution is available via an interface.
•
Impedance and resistance results are calculated from voltage and current measurements. For near zero
and negative current measurements, the 95x displays the maximum meaningful resistance value
preceded by the > character.
REVIEWING TEST RESULTS AFTER RUNNING A TEST SEQUENCE
Immediately after running a test sequence and temporarily displaying an overall pass/fail status message the 95x
automatically initiates reviewing the detailed test results. The results of a previously run test sequence can also be
manually reviewed later by pressing the
VIEW
key if the
READY
indicator is illuminated, indicating that the test
sequence is still ready to be run.
If the test sequence failed, then the review is initiated at the first failed step, otherwise it is initiated at the first
test step in the test sequence.
The following table shows example displays of reviewed results and their meanings. See the descriptions following
the table for a description of each of the fields in the display.
01.01d 1500V 60.0Hz
5.0s PASS 2.10mArms
The measurement being reviewed was within limits.
01.01d 1500V 60.0Hz
5.0s DET 28mAarc
Arcing was detected during the test but the 95x was not configured to fail on arc
detection.
01.01d 1500V 60.0Hz
5.0s FAIL 28mAarc
Arcing was detected during the test which caused the step to fail.
01.01d 1500V 60.0Hz
1.5s >MAX 2.10mArms
The measurement being reviewed was above the maximum limit which caused
the step to fail.
01.01d 1500V 60.0Hz
1.5s <MIN 12.3uArms
The measurement being reviewed was below the minimum limit which caused
the step to fail.
01.01d 1500V
1.5s UNST 12.3uA
The measurement being reviewed was within limits but was unsteady with
increasing leakage current (decreasing resistance) which caused the step to fail.
01.01d 1500V 60.0Hz
HV TRIP EXCEEDED
The HV Trip safety feature was tripped which caused the step to fail. This is
typically caused by a DUT or cable breakdown to ground.