95x Series Operating Manual - May 17, 2022
Page
42
of
155
o
The presence of lead compensation data in a test sequence is indicated on the display during test
sequence selection by the presence of a
c
immediately after the test sequence number.
o
Any saved lead compensation data for a test sequence is automatically discarded if the test
sequence is selected for editing by using the
EDIT
key.
SECTION 6
–
TEST STEPS
The previous section described how to use test sequences. This section describes how to choose and configure
each step within a test sequence and the specifications and limitations for each step type.
The table below gives an overall summary of the various types of activities available for each step. After the table
are individual sections giving full details for each of these activities. The user only need be conversant with the
section for each specific activity which the user intends to perform. The levels shown for each model is those for
the standard build, option content may change the available levels and/or frequencies.
Activity Required
Test Step
Types
951i
952i
953i
954i
955i
956i
957i
959i
AC Voltage
Withstand and
Leakage Testing
ACez
,
ACW
and
ACCAP
20V-6kV,
20-
500Hz
20V-6kV,
20-
500Hz
20V-6kV,
20-
500Hz
20V-6kV,
20-
500Hz
50V-
10kV,
40-
500Hz
No
20V-6kV,
20-
500Hz
No
DC Voltage
Withstand and
Leakage Testing
DCez
,
DCW
and
DCIR
20V-
6.5kV
20V-
6.5kV
40V-
11kV
40V-
11kV
40V-
11kV
20V-
6.5kV
75V-
15kV
No
Pulsed Voltage
Withstand
Testing (Opt.
PMT-1 only)
PULSE
50V-8kV
50V-8kV
50V-8kV
50V-8kV
100V-
10kV
No
50V-8kV
No
DC Voltage
Breakdown
Device Testing
BRKDN
20V-
6.5kV
20V-
6.5kV
40V-
11kV
40V-
11kV
40V-
11kV
20V-
6.5kV
75V-
15kV
No
DC Low
Resistance
Testing
Low
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
1m
Ω
to
150K
Ω
AC Ground Bond
Testing
GBez
and
GB
No
1u
Ω
-
10Ω
@ 0.1-
40A
No
1u
Ω
-
10Ω
@ 0.1-
40A
No
No
Opt.
GB40
only
1u
Ω
-
10Ω
@ 0.1-
40A
Ground Leakage
Testing
ACI
and
DCI
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
0.1nA-
200mA
Switch Unit
Control
SWITCH
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Test Sequence
Timing Control
PAUSE
and
HOLD
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
When the user is selecting testing against the requirements of a safety standard, the information given in this
section should only be used as a guideline; the user should consult a safety testing professional for information
regarding the specific standard and application. The voltage ranges shown above assume the most common
option content for each model; see the specifications for each type of test step for the actual ranges.
CHOOSING WITHIN THE VOLTAGE WITHSTAND AND LEAKAGE TESTING GROUP
When Voltage Withstand and Leakage Testing is required, the user must first decide whether to perform this using
an AC, DC or pulsed voltage, the frequency to use, the limits to use, and the actual test type(s) to perform.