95x Series Operating Manual - May 17, 2022
Page
81
of
155
100-150K
Ω
: 5%
Opt. LRE-10 : add 0.001
Ω
Using rear terminals (option RPO only) : add 0.3
Ω
Lead Compensation
<(±1% of error ± 0.005
Ω
)
MEASUREMENT ACCURACY (4-WIRE)
Measurement Period
7.25msec measurement period (100ms for test time >2 sec)
Accuracy
<30K
Ω
: 0.8% + 0.002
Ω
30-100K
Ω
: 1.5%
100-150K
Ω
: 5%
Opt. LRE-10 : add 0.001
Ω
Lead Compensation
<(±1% of error ± 0.001
Ω
)
4-wire Compensation
SOURCE <(±1% total lead impedance, 100
Ω
max)
SENSE <(±0.1% total lead impedance, 100
Ω
max)
TEST TIMING
Test Time
0.02 to 9999sec or user terminated
<(0.05% + 20ms) accuracy (add 100ms for test time >2 sec)
Delay
0.00 to 9999sec
<(0.05% + 20ms) accuracy (add 100ms for test time >2 sec)
AC GROUND BOND TESTING (GBEZ AND GB)
These types are primarily for testing resistance values from a few microohms up to 8
Ω
at a user set AC test current
between 0.1 and 40Arms, typically for ground bond testing requirements of safety standards and also for high
current connector, cabling, or component testing.
Both GBez and GB are performed in the same manner; the difference between them is in the amount of
configuration available. The GBez type is intended for basic applications, the GB type for comprehensive testing.
The GB type can be configured to provide the same testing as the GBez type.
GBez
GB
User set open circuit voltage limit
No
Yes
RMS Impedance Min/Max Limits
Yes
Optional
In-Phase or Quadrature Impedance Min/Max Limits
No
Optional
Timed Ramp
No
Optional
Timed Discharge
No
Optional
Smooth Transition to next test step if GB type
No
Optional