2000I & 2000M Instrument’s User Guide, Revision B
24
When the POWER
switch is placed in the OFF
position, the terminals remain in
the condition prevailing.
Vitrek
recommends selecting the “Stand-by” state
prior to switching the POWER
switch OFF, to afford the highest level of
protection both to any circuitry connected to the instrument and to protect the
instrument from inadvertent application of external signals.
Connecting to the OUTPUT Terminals
To deliver voltage, current, or temperature to your test equipment, connect it to
the binding-post terminals using banana plugs, spade leads or wire. When a posi-
tive polarity of output is requested, the RED terminal is the positive terminal and
has the requested current flow.
Notes:
1)
The terminals are automatically configured as input terminals when the
2000M instrument is set to the “Measure”
mode.
2)
In cold junction compensation mode (CJC), the output connectors are wired
in parallel with a female rear-panel mounted connector.
Displaying Instrument Status
The DC calibrator instrument offers the flexibility of setting the output level and
setting the instrument’s status to “Standby” or “Operate”
.
This is done following
the application of power. The display shows the output level and status following
the messages associated with the power on internal tests. With the instrument in
the user-selectable “Standby” mode, the output connectors are disconnected in-
ternally.
The following example is seen in “Operate” status, which means the output level
is actually being produced at the output terminals of the instrument. Examples of
the output level and status displays are:
< +10.00000mV >
The following example is seen in “Standby” status, which means the output ter-
minals are open circuited internally.
<+ 10.00000mV Stby >
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...