2000I & 2000M Instrument’s User Guide, Revision B
30
2
.
Press the
+/-
key to change the simulated temperature to:
-100.00ºC R
Note that unlike voltage or current output levels, the limits of available simulation
temperatures are not identical in both polarities (in general). If the attempted tem-
perature exceeds these limits, then the simulation temperature will be set to the
limit at the requested polarity and a warning message will be displayed for ap-
proximately two seconds.
Adjusting Output Level or Simulated Temperature
The 2000 instruments allow you to incrementally step up, or down, the output
level and to incrementally step up, or down, the hot junction simulation tempera-
ture. Use this “Adjust” mode to perform calibration of an external unit of equip-
ment. During a calibration procedure the 2000 instrument will display each
modified output level of the equipment being calibrated. This function allows for
any error in the equipment to be highlighted. This section describes the method
used to perform this action.
To access “Adjust” mode:
1.
Press the MNU key and then an arrow
key: 2, 4, 6, or 8 repetitively until the
adjust choice displays:
< Start Adjust >
2.
Press any of the Units/Enter keys (V, A, ºC or ºF
)
to select the “Adjust”
mode. The display then returns to displaying the selected output level, with
the least significant displayed digit flashing (showing the size of the incre-
ment that will be used).
During the “Adjust” mode of operation, the following changes to the display for-
mat occur:
The display no longer shows the “Stby” characters at the right hand end of
the display when in “Standby” mode.
The multiplier is automatically selected to provide the required resolution
in the display. The previously described ability to control the multiplier
used in the display format is inactive.
The thermocouple type data is no longer provided in the display.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...