2000I & 2000M Instrument’s User Guide, Revision B
46
with each instrument as standard and when MEM is ordered. (
Vitrek
part
number
C109001001000001).
General Memory Considerations
Memory of Instument Status
The contents of a memory step only affect the selection of the “Operate” or
“Standby” status when the memory step is initially recalled. To change the state
following recall use the method outlined in
Front Panel Operation
.
Editing Memory Steps
The contents of a memory step may be edited at any time, if the memory step
being edited is presently in use, then note that the present output level will be
updated when the memory step is next recalled.
Adjusting the Output Level
The “Adjust” mode
of operation of the instrument may be used in conjunction
with recalling memory steps. Recalling a memory step while in this mode
sets the
actual output level and the nominal output level to that recalled from the memory.
In this manner you are not required to deselect “Adjust” mode when sequentially
recalling steps from memory.
Thermocouple Simulation
Only the simulation temperature and the temperature units are stored in a memory
location. This feature allows you to change thermocouple type or change between
Active
and
Passive Cold Junction Compensation
without reprogramming the
memory steps.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...