Applications 55
the required correction, in the units used by the transducer. Note that this correc-
tion may be of either polarity.
Note:
The entered data are non-volatile, thus they may be entered at your base
location, and then be used at a remote site without the need to reenter these data.
Refer to
Using the Memory Options
.
Adjusting Current Measuring of Multiple Units
The 2000 instruments may be used to adjust multiple-current measuring equip-
ment simultaneously. This is achieved by connecting the units in series. Here the
18V compliance capacity of the 2000 instrument allows for as many as 15 units
to be adjusted simultaneously. When connecting line-powered current measuring
equipment in series check for excessive leakage-current flowing from an input
terminal to ground. Current leakage does not affect the measurement when a sin-
gle unit is connected, but each unit will affect the indications of others when con-
nected in series.
Also many highly-sensitive current measuring devices have inferior rejection to
interference on one of their input terminals. When connected in series the interfe-
rence generated by one unit can adversely affect the performance of the next. To
operate multiple units in series, test for loss of accuracy during this procedure.
Before adjusting any equipment check the calibration of each unit individually,
then connect the equipment in series and recheck.
Note:
The low output impedance of the 220mV and 22mV ranges of the 2000
instruments enable a small correction factor to be easily applied when the instru-
ment is driving low load impedances.
Simulating Thermocouple Systems
2000M and 2000MN instruments can be used to simulate single and dual junction
thermocouple systems. This instrument offers the choice of using the internal
cold junction
compensation or using an external high accuracy ice-point reference
junction. When using internal cold junction
compensation to achieve thermo-
couple simulation,
the
accuracies approach 0.1ºC. When using external ice-point
reference junction to achieve thermocouple simulation, accuracies approach
0.02ºC
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...