2000I & 2000M Instrument’s User Guide, Revision B
58
The 2000 instruments have an output impedance greater than 10
11
, thus this
affect is not of importance for unknown values below 1M
. The following table
gives the uncertainty of the transfer for various values of standard and unknown
resistors. See Note following table.
Table 1: Resistance Values for Transfer
STANDARD
UNKNOWN
CURRENT
BASIC UNCER-
TAINTY
UNCERTAINTY
WITH 1PPM + 1
μ
V
DVM STABILITY
100
10
20mA 5.75ppm
11.25ppm
10
100
20mA 5.75ppm
7.25ppm
1K
100
10mA 6.5ppm
8.5ppm
1K
1K
10mA 6.5ppm
7.6ppm
10K
1K
1mA 6.1ppm
8.1ppm
10K
10K
1mA 6.0ppm
7.1ppm
100K
10K
100
μ
A 7.5ppm
9.5ppm
100K
100K
100
μ
A 6.5ppm
7.6ppm
1M
100K
10
μ
A 20ppm
22ppm
1M
1M
10
μ
A 10ppm
11.1ppm
10M
1M
1
μ
A 137ppm
139ppm
10M
10M
1
μ
A 37ppm 38.1ppm
100M
10M
0.1
μ
A 1300ppm
1300ppm
100M
100M
0.1
μ
A 307ppm
308ppm
Note:
The
Basic Uncertainty
figures in this table are those associated with the
instrument’s stability only, while the overall uncertainty figures are those includ-
ing 1ppm + 1
V uncertainty for the voltmeter, which can be achieved with a high
precision DVM, e.g., Hewlett-Packard 3458A. Note also that, in a standards
la-
boratory environment, up to 2:1 better uncertainties are readily attainable from
the 2000 instruments.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...