Calibration and Maintenance
65
Note:
To abort the external calibration selection at any time, press the CLR
key. You will return the instrument to the output level and status that was
present when the MNU key was initially pressed.
6.
In no password is requested or the correct password is entered, press any of
the arrow keys to continue.
7.
Whether a password was previously required or not, you may be prompted to
enter a new password:
< New Password >
To bypass entering a new password, press any of the arrow keys to
continue.
To enter a password, press any of the Units/Enter keys (V, A, ºC or
ºF). The following will display:
<Global Password?>
To enter a password at this time, press the digits for the new password
(up to 10 digits). Press the CLR key to abort, as required. Press any of
the Units/Enter keys to place the password in non-volatile storage. Use
this entered password for all subsequent external calibrations. If the
user presses a Units/Enter key (V, A, ºC or ºF), having pressed no nu-
meric keys, then a null password is entered, which effectively disables
the password protection system.
If the previously selected password has been lost, use the common
password for all 2000 instruments, which is listed on the last page of
this user guide. Note that this page/password may have been filed to
prevent access. The following will display:
<Select password>
8.
The instrument is ready to begin the external calibration sequence.
Performing External Calibration
The instrument automatically selects the next output level and selects the “Ad-
just” mode of operation, with the least significant digit as the adjustment digit.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...