2000I & 2000M Instrument’s User Guide, Revision B
66
Front Panel Key Adjustments
The
left and
right arrow
keys select the adjustment digit, the
up and
down arrow
keys adjust the selected digit. To abort any external calibration step
press the CLR key. To abort the calibration entirely press the MNU key. To nor-
mally end a calibration step—thus entering the adjusted calibration factor, press
any of the Units/Enter keys.
Note:
Pressing a Units/Enter key while the displayed deviation is at zero will
effectively skip the procedure to the next calibration
point without affecting the
calibration factor.
Important Note:
Take precautions against thermally generated voltages during
the execution of the external calibration as extremely low-level DC voltage mea-
surements are required. Be advised of the
Applications
chapter
discussion
regard-
ing low-level DC voltage connections.
1.
Attach the 2000I or 2000M instrument to a recently zeroed nullmeter.
2.
Measure the actual output level of the 2000 instrument using the nullmeter.
Adjust the instrument, using the arrow
keys, until the output level is measured
to
0.0
V
, within
0.1
V, 0.2
V, 1
V
and
10
V
respectively.
3.
Calibrate to an output level of
+10.00000mV
. Adjust the output level of the
2000 instrument to within
0.2
V
of
10mV
as described in step 2.
2000
Nullmeter
2000
Nullmeter
100:1 Divider
1V DC
Source
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...