Calibration and Maintenance
67
This output level may be measured directly by using:
A high precision DMM, or;
A nullmeter to monitor the deviation between this instruments' output
level and that of a precision calibrated
10mV
source, or;
The preferred method of using a nullmeter to monitor the deviation be-
tween this instrument’s output level and the output of a high precision
100:1
attenuator whose input is obtained from a precision calibrated
1V
source.
4.
Calibrate to an output level of
+100.000mV
. Adjust the output level to with-
in
0.7
V
of
100mV
as described in step 2.
This output level may be measured directly by using:
A high precision DMM, or;
A nullmeter to monitor the deviation between this instrument’s output
level and that of a precision calibrated
100mV
source, or;
The preferred method of using a nullmeter to monitor the deviation be-
tween this instrument’s output level and the output of a high precision
100:1
attenuator whose input is obtained from a precision calibrated
10V
source.
5.
Calibrate to an output level of
+1.000000V
. Adjust the output level of the
instrument to within
6
V
of
1V
as described in step 2.
2000
Nullmeter
100:1 Divider
10V DC
Source
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...