2000I & 2000M Instrument’s User Guide, Revision B
68
This output level may be measured directly by using:
A high precision DMM, or;
A nullmeter to monitor the deviation between this instrument’s output
level and that of a precision calibrated
1V
source, or;
The preferred method of using a nullmeter to monitor the deviation be-
tween this instrument’s output level and the output of a high precision
10:1
attenuator whose input is obtained from a precision calibrated
10V
source.
6.
Calibrate to an output level of
+10.00000V
. Adjust the output level of the
2000 instrument to within
60
V
of
10V
as described in step 2.
This output level may be measured directly by using:
A high precision DMM, or;
The preferred method of using a nullmeter to monitor the deviation be-
tween this instrument’s output level and that of a precision calibrated
10V
source.
7.
Calibrate to an output level of
-10.00000V
. Adjust the output level of the
2000 instrument to within
60
V
of
10V
as described in step 2.
2000
Nullmeter
10:1 Divider
10V DC
Source
2000
Nullmeter
10V DC
Source
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...