Calibration and Maintenance
71
This output level may be measured directly by using a calibration standard
1K
resistor and a high precision DMM to monitor the voltage drop across
the standard resistor.
15. Calibrate to an output level of
+10.00000mA
. Adjust the output of the 2000
instrument as described in step 2, until the voltage measured is
1.0V +/–
20
V
(assuming exact resistance of 100
), i.e., a current of
10mA +/–
200nA
.
This output level may be measured directly by using a calibration standard
100
resistor and a high precision DMM to monitor the voltage drop across
the standard resistor.
16. The next calibration point will be to measure
zero (0.0)
VDC
. Attach a small
piece of #24 AWG buss wire to the output terminals and allow at least two
minutes for the reading to settle.
17. The final calibration point will be to measure
0.1VDC
. Remove the buss wire
and apply
0.1VDC
to the output terminals. Allow at least two minutes for set-
tling.
18. No further calibration steps are required. The instrument will now store the
calibration factors obtained during this procedure and then select the “Stand-
by”
state with the power on default output level.
Note:
Laboratory standard resistors are not normally the exact nominal value.
The actual measured voltage drop used for each DC current calibration step will
require adjustment for the actual value of the resistor.
V
MEASURED
= I
REQUIRED
*
R
ACTUAL
Periodic Maintenance
The 2000I(N) and 2000M(N) DC power source instruments require little
periodic maintenance; however,
Vitrek
makes the following recommenda-tions:
1. Allow the internal battery to fully discharge, and then fully recharge at least
once a year. This will ensure that the battery retains its storage capacity.
2. Clean the external surfaces of the chassis with a soft damp cloth. Do not clean
the front panel with Freon based cleaning solutions. Take care that moisture
does not enter the chassis during cleaning. If the external surfaces become ex-
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...