2000I & 2000M Instrument’s User Guide, Revision B
72
cessively dirty or damaged, contact
Vitrek
Customer Service (or one of its
agents) for details on how to rectify this situation.
3. If the instrument is to be stored for extended periods of time, fully charge
before storing. This will ensure that the battery will accept a full charge when
removed from storage.
Internal Maintenance
In general, the 2000I and 2000M DC power source instruments do not require
internal maintenance. In the event of a failure, a calibration fault, or when the
internal battery requires replacement, however, the instrument may require open-
ing.
Important Note:
The information provided here allows you to replace the bat-
tery. If you wish to perform detailed fault finding within this instrument, obtain
the MS2000 option
Vitrek
.
Caution:
The 2000 instruments contain delicate precision circuitry; ensure static
control procedures are followed while internal maintenance is in progress.
Warning:
Prior to removing the outside covers of the 2000 instruments, ensure
that the external charging unit or power supply is disconnected and check that the
front panel’s POWER
switch is OFF.
To obtain access to the instrument’s internal circuitry:
Note:
Ensure static control procedures are followed during disassembly and
reassembly.
1. Remove the
four
lower feet by pulling them horizontally away from the chas-
sis.
2. Remove the
two
cross-head screws exposed by the removal of the feet in
step 1.
3. Remove the
two
cross-head screws holding the top of the two halves of the
outer chassis covers.
4. Carefully remove
both halves
of the outer chassis covers by pulling them
horizontally out and away from the internal chassis.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...