2000I & 2000M Instrument’s User Guide, Revision B
74
To replace a battery:
1.
Open the instrument and remove the Microprocessor board as described in
the instruction:
To obtain access to the internal circuitry
. Additionally re-
move the Analog board and loosen the Power Supply board, as required, to
assist in replacing the battery.
2.
Remove the
two
#6 nuts and corresponding screws that hold the battery strap
to the bracket. (Instruments with studs, require only removal of the nuts.)
3.
Remove the battery strap and old battery.
4.
Install a new battery and reassemble the instrument in reverse order of disas-
sembly instructions. Be careful to replace all cable interconnections, using
the same orientation and cable routing as was present prior to disassembly.
Ensure the nylon washers are included when reattaching the Analog board.
Warning:
The printed circuit boards are continuously powered by the battery,
use extreme care in handling them. Reconnect the battery cables after reattach-
ment of the boards.
5.
Following complete reassembly, fully charge the replaced battery, in accor-
dance with the charging procedures in the
Battery Charging
chapter.
6.
Perform an internal and external recalibration on instrument after charging in
accordance with the calibration procedures.
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...