Interface Operation
81
Char/Seq
Description of Action
O
Selects the “Operate”
state.
P
Commands the interface to assert the
SRQ
(service request) line
when a previously received command has been fully actioned. Ap-
plies to UE2000 only.
R
Commands a
power-on
reset sequence.
S
Selects the “Standby”
state.
T
Sets the Thermocouple type to that described by the character im-
mediately following, e.g., “TK”
sets the thermocouple type to “K”
.
X
Initiates an external calibration, which is equal to selecting “Exter-
nal Cal”.
Firmware Version F.02 or Newer
The Firmware Version F.02 update to the 2000I, 2000M, 2000IN and 2000MN
DC power source instrument models include all the standard capabilities, includ-
ing “Range Lock” mode, the ITS-90 Thermocouple temperature for the 2000M
and 2000MN instruments and an extended command set.
Extended Command Set
The interface command list includes “Range Lock” and “Measurement” modes.
The software version F.02 command set for the 2000 instruments and RS-BAT
and IE-BAT interfaces is given in Table 6.
Table 6: Extended Command Set
Command
Description
G0
Deselects the “Measure” mode
G1
Selects the “Measure” mode
L1
Locks the 20 volt Range
L2
Locks the 2 volt Range
L5
Locks the 20mA Range
Tx
Selects thermocouple type x (2000M only)
UA
Unlocks all ranges: voltage and current
U1
Unlocks the 20 volt Range
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...