Appendix
A—Output
Specifications 85
Appendix A–
Output Specifications
The DC voltage and current output capabilities of the 2000I,2000M, 2000IN and
the 2000MN instruments are provided here. Additionally, specifications for the
2000M instrument’s of thermocouple measurement and simulation option are
included here.
DC Voltage Output Specifications
The accuracy figures given in Table 7 are valid throughout the temperature range
of 18º to 28ºC. The 24 hour stability figures are for constant load and temperature
(within 1ºC) and are measured using a DC to 2Hz bandwidth. The accuracy fig-
ures are relative to calibration standards. Add 3ppm + 1µV to the voltage figures,
and 15ppm + 20pA to the current figures, for absolute accuracy.
Table 7: DC Voltage Output
Levels
<22mV
22 - 220mV
0.22 - 2.2V
2.2 - 22V
Resolution 10nV
100nV
1
V 10
V
24 Hr. Stability (ppm +
V)
3.0 + 0.08
2.5 + 0.1
1.8 + 0.5
1.8 + 3.0
90 Day Accuracy (ppm +
V
)
20 + 0.2
15 + 0.25
12 + 1.0
10 + 4.0
1 Year Accuracy (ppm +
V)
30 + 0.2
25 + 0.25
20 + 10.0
18 + 4.0
Temperature Coefficient P
(ppm +
V)
2.5 + 0.03
2.0 + 0.04
1.5 + 0.2
1.5 + 1.0
Pk-PK Noise 0.01 – 10 Hz
(ppm+
V/
C)
0.3 + 0.04
0.3 + 0.06
0.3 + 0.3
0.3 + 2.0
RMS Noise 10Hz – 10KHz
5µV
5µV
7.5µV
10µV
Summary of Contents for Xitron 2000 Series
Page 1: ...USER GUIDE 2000 Family DC Temperature Source Instruments...
Page 2: ......
Page 18: ...2000I 2000M Instrument s User Guide Revision B 18...
Page 42: ...2000I 2000M Instrument s User Guide Revision B 42...
Page 47: ...Using the Memory Options 47...
Page 60: ...2000I 2000M Instrument s User Guide Revision B 60...
Page 91: ......
Page 92: ...2000I 2000M Instrument s User Guide Revision B 92...
Page 95: ...Appendix B General Specifications 95...
Page 97: ...Password Page 97...