22
2553/ 2553E Command Guide
CLR-INTEGRATE
This command clears all integrated measurements to zero.
No other measurement results are affected by this command.
This command is implied whenever measurements are started using the MEASURE
command.
Resetting the DC Current Offsets
SET-DC-ZERO
This command sets the DC current zero offsets to the presently measured values for all
three phases.
Both the DC and RMS values of the measured current must be less than 2% of the full-
scale value for the 2553/ 2553E on all three phases, otherwise a command error is
generated.
Only in rare situations will this command be used when the 2553/ 2553E has frozen
measurement results.
This command may be issued when the 2553/ 2553E is configured for AC-ONLY with
no apparent effect since only the DC offsets will be updated.
Setting Default Measurement Configuration
SETDEFAULTS
This command sets the measurement configuration to factory default values. Use this
command to ensure that a 2553/ 2553E is placed into an exactly known state. The action
of this command is equivalent to the following commands.
•
BANDWIDTH=1
Measurement bandwidth is set to 20Hz to 5KHz
•
WIRING=3P4W
2553/ 2553E is wired for 3-phase 4-wire operation.
•
AC-ONLY=0
Measurements include both AC and DC content
•
AVERAGE=1
Measurements are averaged over 250ms
•
SYNC=0
Measurements are synchronized to the phase A voltage signal
•
MEASURE=1
Normal measurements are enabled
•
INTEGRATE=0
Integrated measurements are stopped, and their results are
cleared
•
HISTORY=1
Historical results storage is enabled, and any stored results are
cleared
•
CLR-INRUSH
Inrush results are cleared
•
CLR-INTEGRATE
Integrated results are cleared
It is not recommended to issue other commands in the same set as this command.