Sending Commands to the 2553/ 2553E
25
BANDWIDTH
This command sets the bandwidth to be used for standard measurement results. The
BANDWIDTH command takes a single ASCII character as data.
Character
Bandwidth
0
20Hz to 75KHz
1
20Hz to 5KHz
2
2Hz to 2KHz
3
0.2Hz to 200Hz
4
0.02Hz to 20Hz
Table 4 – Data for the BANDWIDTH command
Notes –
•
The BANDWIDTH? Interrogative command returns the present state of the
BANDWIDTH command.
•
If measurements are not frozen when this command is executed, then all
measurements will be restarted as if the MEASURE=START command had also
been executed at the same time.
•
The SETDEFAULTS command implies BANDWIDTH=1.
Example –
•
The user transmits: BANDWIDTH=1<NL>
The 2553/ 2553E selects a 20Hz to 5KHz bandwidth for measurement results, and
will restart all measurements if the 2553/ 2553E is presently making measurements.
HISTORY
This command is used to set whether the historical data is to be updated or frozen. The
HISTORY command takes a single data only, which may be any of the following ASCII
data.
ASCII Data
Command Action
0
Historical data is frozen
STOP
Historical data is frozen
1
Historical data is updated at the commanded rate
START
Historical data is updated at the commanded rate
Table 5 – Data for the HISTORY command