26
2553/ 2553E Command Guide
Notes –
•
HISTORY=1 or HISTORY=START clears any presently stored historical data, even
if historical data storage was already enabled.
•
If all measurement results are frozen when the HISTORY=1 (or
HISTORY=START) command is received, then both standard and historical
measurements will be started.
•
Any commands that affect starting, stopping or restarting measurements also affect
historical data.
•
Although there is a slight increase in product speed when the history data collection
process is disabled, this would rarely yield any benefit thus would be rarely used.
Normally the MEASURE command would be used to start and stop all
measurements, including the historical results.
•
The HISTORY? Interrogative command returns the present state of the HISTORY
command.
•
See the HISTORY-SCALE command for setting the rate at which the historical data
is updated.
Example –
•
The user transmits: HISTORY=STOP<NL>
The 2553/ 2553E will stop updating the historical results, leaving all results frozen
with the present contents. If the historical results were already frozen, then this has
no affect.