Sending Commands to the 2553/ 2553E
27
HISTORY-SCALE
This command sets the rate at which historical data is stored by the 2553/ 2553E. Any
presently stored historical data is cleared when this command is actioned. The
HISTORY-SCALE command takes a single data only, which may be any of the
following ASCII data.
See the HISTORY command for starting and stopping the update of historical data.
ASCII Data
History Update Rate
0
0.4 seconds per division
1
1 second per division
2
2 seconds per division
3
5 seconds per division
4
10 seconds per division
5
30 seconds per division
6
1 minute per division
7
3 minutes per division
8
10 minutes per division
9
30 minutes per division
10
1 hour per division
11
3 hours per division
12
6 hours per division
13
12 hours per division
14
1 day per division
Table 6 – Data for the HISTORY-SCALE command
Notes –
•
When enabled, all historical data is updated.
•
Historical data is always cleared by this command.
•
There are 20 equally spaced historical data points per division, thus the time between
each stored data point is 1/20
th
of the times indicated above.
•
All peak data are stored as the highest peak recorded between each data point.
•
All other (non-peak) data are stored as the average value recorded between each data
point. Historical data is unaffected by the selected measurement result averaging
period.
•
Any commands that affect starting, stopping or restarting measurements also affect
historical data.
•
The HISTORY-SCALE? Interrogative command returns the present state of the
HISTORY-SCALE command.