40
2553/ 2553E Command Guide
Example –
•
The user transmits: *CAL?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>0<NL>
This indicates that the 2553/ 2553E was successfully calibrated.
CAL-DATE?
The 2553/ 2553E places an ASCII string indicating the date on which the 2553/ 2553E
was last successfully calibrated against external standards. The string NOT
CALIBRATED is returned if the 2553/ 2553E has not been successfully calibrated, or the
previously stored calibration has become corrupted.
Example –
•
The user transmits: CAL-DATE?<NL>
The user reads the 2553/ 2553E and receives the characters:
<SP>Apr<SP>28<SP>1998<NL>
This indicates that the 2553/ 2553E was last successfully calibrated on the 28
th
of
April, 1998.
Interrogating Printer Status
PRINT-STATUS?
The 2553/ 2553E places a single ASCII character indicating the status of the parallel
printer port of the 2553/ 2553E. Note that the user may only command a printout (either
manually or by means of the PRINT command) when the printer port is presently idle. If
multiple printouts are required, then the user must poll this interrogative until the port is
idle prior to transmitting another PRINT command. The character returned is one of the
following.
Character
Printer Status
0
Printer interface is idle
1
Printer interface is active, with no errors
2
Printer interface is active, with a paper out indication
3
Printer interface is active, with an offline indication
4
Printer interface is active, with a printer error indication
Table 23 – Data returned due to PRINT-STATUS? command
Summary of Contents for Xitron 2553
Page 1: ...IEEE488 COMMAND GUIDE 2553 2553E Three Phase Power Analyser ...
Page 2: ......
Page 10: ...10 2553 2553E Command Guide ...
Page 38: ...38 2553 2553E Command Guide ...