46
2553/ 2553E Command Guide
HISTORY-SCALE?
The 2553/ 2553E places a pair of ASCII characters representing the presently selected
timescale for historical measurement results. The HISTORY-SCALE command controls
this configuration. The characters returned are as follows.
Characters
History Timescale
<SP>0
0.4 seconds per division
<SP>1
1 second per division
<SP>2
2 seconds per division
<SP>3
5 seconds per division
<SP>4
10 seconds per division
<SP>5
30 seconds per division
<SP>6
1 minute per division
<SP>7
3 minutes per division
<SP>8
10 minutes per division
<SP>9
30 minutes per division
10
1 hour per division
11
3 hours per division
12
6 hours per division
13
12 hours per division
14
1 day per division
Table 28 – Data returned due to HISTORY-SCALE? command
Example –
•
The user transmits: HISTORY-SCALE?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP><SP>1<NL>
This indicates that the 2553/ 2553E is storing historical measurement results at the
rate of 1 second per division.
INTEGRATE?
The 2553/ 2553E places the ASCII character 1 (decimal 49) if it is presently integrating
measurements, or the ASCII character 0 (decimal 48) if not. Note that the 2553/ 2553E is
not updating integrated results whenever measurements are frozen. (See the
INTEGRATE and MEASURE commands.)