Reading Measurement Results from the 2553/ 2553E
63
VAR[
phase
/
harmonic
-
harmonic
]
The imaginary power measurement result evaluated for a range of harmonics. The result
format is floating point, in VAR.
VAR-INRUSH[
phase
/RMS]
The highest inrush imaginary power measurement result, including DC if configured (see
the AC-ONLY command), over the full configured bandwidth (see the BANDWIDTH
command), including any non-harmonic content. The result format is floating point, in
VAR. VAR-INRUSH[RMS] may also be used for this definition.
WATTS[
phase
/DC]
The same as VA[DC]. If configured as AC only then this result is always zero. (See the
AC-ONLY command). The result format is floating point, in Watts.
WATTS[
phase
/FUND]
The real power measurement result evaluated for the fundamental component only. The
result format is floating point, in Watts.
WATTS[
phase
/RMS]
The real power measurement result, including DC if configured (see the AC-ONLY
command), over the full configured bandwidth (see the BANDWIDTH command),
including any non-harmonic content. The result format is floating point, in Watts.
WATTS[
phase
/
harmonic
]
The real power measurement result evaluated at the specified harmonic. The result format
is floating point, in Watts.
WATTS[
phase
/
harmonic
:
harmonic
]
The real power measurement result evaluated for each harmonic. Each result format is
floating point, in Watts.
WATTS[
phase
/
harmonic
-
harmonic
]
The real power measurement result evaluated for a range of harmonics. The result format
is floating point, in Watts.