Frequently Asked Questions
83
Frequently Asked Questions
This section provides some answers to the most commonly asked questions regarding
interfacing the 2553/ 2553E to an IEEE488 controller.
Vitrek
is
continually updating
its knowledge of the answers to questions asked by customers. If
your question is not
listed here, or the answers given do not apply to your particular
situation, please
contact
Vitrek.
The controller issues a message that there is no listeners, but the
2553/ 2553E is connected. What is wrong?
This can be caused by a great many problems. The most common issues are –
1.
A faulty IEEE488 cable, or the cable is not correctly mounted on the controller
and/or the 2553/ 2553E. Carefully visually check the contacts on the connectors of
the cable, the controller and the 2553/ 2553E.
2.
Check the address setting for both the controller and the 2553/ 2553E. These must
not be the same address.
3.
Check if there are multiple applications using the same IEEE488 controller. In some
circumstances, and under some operating systems, this cannot be performed due to
controller or software limitations.
4.
Attempt to communicate with (i.e. send commands to) the 2553/ 2553E. Use a
simple command like MEASURE=STOP. Under some circumstances, when the
2553/ 2553E is first plugged into the IEEE488 cable, it has been found that some
controllers do not clear this error automatically when the error does not exist
anymore. This can occur when the 2553/ 2553E is first plugged into the bus, or after
a device clear bus command.
5.
Reboot the controller and the 2553/ 2553E with the IEEE488 cable attached. Under
some circumstances it has been found the certain controllers can achieve an illegal
handshake condition which cannot be recovered from when the cable is first
attached.
6.
Check that the controller has not been configured to perform any “special”
handshake techniques. The 2553/ 2553E does not utilize special "high speed"
handshaking, such as the National Instruments HS488 technique.