44
2553/ 2553E Command Guide
Example –
•
The user transmits: STATUS?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP><SP>20<NL>
This indicates that the 2553/ 2553E has new data available to be read (always the
case for this command), and that the voltage input has been overloaded at some time
since the status byte was last cleared.
Interrogating the Measurement Configuration
AC-ONLY?
The 2553/ 2553E places the ASCII character 1 (decimal 49) if it is presently making
measurements without the DC content, or the ASCII character 0 (decimal 48) if the
measurements include the DC content. The AC-ONLY command controls this
configuration.
Example –
•
The user transmits: AC-ONLY?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>0<NL>
This indicates that the 2553/ 2553E is making measurements which include the DC
content of the signals.
AVERAGE?
The 2553/ 2553E places a single ASCII character representing the presently selected
averaging period for measurement results. The AVERAGE command controls this
configuration. The characters returned are as follows.
Character
Averaging Period
0
50ms
1
250ms
2
1s
3
2.5s
4
5s
5
10s
6
20s
7
1 minute
Table 26 – Data returned due to AVERAGE? command