Reading Status & Settings from the 2553/ 2553E
45
Example –
•
The user transmits: AVERAGE?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>1<NL>
This indicates that the 2553/ 2553E is making measurements averaged over a period
of 250ms.
BANDWIDTH?
The 2553/ 2553E places a single ASCII character representing the presently selected
bandwidth for measurement results. The BANDWIDTH command controls this
configuration. The characters returned are as follows.
Character
Bandwidth
0
20Hz to 100KHz
1
20Hz to 5KHz
2
2Hz to 2KHz
3
0.2Hz to 200Hz
4
0.02Hz to 20Hz
Table 27 – Data returned due to BANDWIDTH? command
Example –
•
The user transmits: BANDWIDTH?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>1<NL>
This indicates that the 2553/ 2553E is making measurements using a bandwidth of
20Hz to 5KHz.
HISTORY?
The 2553/ 2553E places the ASCII character 1 (decimal 49) if it is presently storing
historical measurements, or the ASCII character 0 (decimal 48) if not. Note that the
2553/ 2553E is not updating historical results whenever measurements are frozen. (See
the HISTORY and MEASURE commands.)
Example –
•
The user transmits: HISTORY?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>1<NL>
This indicates that the 2553/ 2553E is storing historical measurement results.