Sending Commands and Interrogatives 119
MINFUND=value – sets that the highest fundamental frequency to which the
2801/2 will synchronize is that given by value (in Hz).
MINFUND? – queries the present MINFUND setting
MAXFUND –
Maximum Fundamental Frequency
This is only effective when MSRTYPE=4
MAXFUND=value – sets that the lowest fundamental frequency to which the
2801/2 will synchronize is that given by value (in Hz).
MAXFUND? – queries the present MAXFUND setting
MAXHARMS –
Maximum Number of Measured Harmonics
This is only effective when MSRTYPE=4
MAXHARMS=value – sets that harmonics up to that set by value (integer)
will be measured. MAXHARMS=0.disables harmonics measurements (and
waveform collection and capture).
MAXHARMS? – queries the present MAXHARMS setting
GENDNRANGEDLY –
User Set Minimum Downrange Delay
This is only effective when MSRTYPE=4
GENDNRANGEDLY=value – sets that the minimum delay before
downranging is set by value (in seconds)..
GENDNRANGEDLY? – queries the present GENDNRANGEDLY setting
MSRDCRIPPLE –
DC Ripple Measurement Method
This is only effective when MSRTYPE=3
MSRDCRIPPLE=0 – sets that DC ripple measurements will be wideband AC
measurements (i.e. non-frequency specific).
MSRDCRIPPLE=1 – sets that DC ripple measurements will be made at
integer multiples of a user specified frequency (see RIPPLEFREQ below).
MSRDCRIPPLE? – queries the present MSRDCRIPPLE settings.
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...