Using the Power Analyzer
45
•
Optionally, set
Wiring Impedance
to (W1 + W2) ohms. Include all
wiring resistances between the source and the points at which the
voltage is measured in the entered value for wiring impedance.
•
Set V Burden Compensation to YES.
•
If the current measurement terminals are reversed, then set
Reverse
Current
to
YES
, otherwise set to
NO
.
•
If the voltage measurement terminals are reversed, then set
Reverse
Voltage
to
YES
, otherwise set to
NO
.
Measurement: 3P-3W 2 Meter Connection, Method 1
Source
Source
V
-
+
Load
Channel 1
Channel 2
Load
V
+
-
+
A
-
+
-
A
Source
Load
X
Z
Y
X
Z
Y
Figure 22. Measure 3P-3W 2 Meter Connection, Method 1