Using the Power Analyzer
57
RESULTS AVERAGING:
MEASUREMENT TYPE:
RANGING AND SCALING …
WIRING ADJUSTMENTS …
LINE SWITCH & INRUSH DETECTION …
WAVEFORM CAPTURE …
RESULTS AVERAGING
All displayed measurements can be averaged using a two-pole filter having a
user-selected time constant. This setting only affects the displayed
measurement results, it does not affect any charted measurement results.
•
NONE; 10ms; 20ms; 50ms;
100ms
; 200ms; 500ms; 1 Sec; 2 Sec; 5
Sec; 10 Sec; 20 Sec; 1 Min
MEASUREMENT TYPE
This menu item contains the following seven setup selections –
LINE FREQ (<1kHz)
Configures the analyzer to make optimal signal measurements at line
frequencies in the 40 to 1000Hz range. To accomplish this, the analyzer –
1) Synchronizes on the voltage waveform with frequencies in the range
of 40Hz to 1000Hz; and
2) Tracks the input signals using a low-pass filter where the filter corner
frequency automatically adjusts to maintain better than 5% accuracy at
200 times the fundamental frequency.
Note:
When you enable autorange and a downrange is required, the
analyzer downranges at the rate of 20ms per range. LINE FREQ (40-
1000Hz) is the recommended selection when making line measurements.
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...