Using the Power Analyzer
59
cycle of the entered base ripple frequency. All harmonic data is measured
over the longer of 5ms (rounded upwards to the nearest integer number of
cycles) or four cycles of the entered base ripple frequency. Frequency is
not measured. Signals have a fixed-corner, low-pass filter applied with a
corner frequency equal to 200 times the entered base ripple frequency
Note:
When you enable autorange and a downrange is required, the
analyzer downranges at the rate of 20ms per range. DC is the
recommended selection when making DC measurements.
USER SET
This selection brings up a menu choice in the Ch1 MEASUREMENT
CONFIGURATION that allows you to directly select USER SET
MEASUREMENT configuration parameters, which are automatically set
for the other selections. Refer to USER SET MEASUREMENT below.
Note:
This USER SET selection is not recommended for inexperienced
users.
LOW FREQ (0.5-20Hz)
Configures the analyzer to make optimal signal measurements at
frequencies less than line frequency, in the range 0.5 to 20Hz. To
accomplish this, the analyzer –
1) Synchronizes on the voltage waveform with frequencies in the range
of 0.5Hz to 20Hz; and
2) Tracks the input signals using a fixed low-pass filter where the filter
corner frequency is set to maintain better than 5% accuracy at 10kHz.
Note:
When you enable autoranging and a downrange is required, the
analyzer downranges at the rate of 10s per range. The LOW FREQ (0.1-
100Hz) selection is recommended when making low frequency
measurements. LOW FREQ is not recommended for line measurements.
LOW FREQ (0.02-5Hz)
Configures the analyzer to make optimal signal measurements at
frequencies less than line frequency, in the range 0.02 to 5Hz. To
accomplish this, the analyzer –
1) Synchronizes on the voltage waveform with frequencies in the range
of 0.02Hz to 5Hz; and
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...