64
2801/2802 Power Analyzer User Guide
from the analyzer.
LOAD
or SOURCE
o
WIRING COMPENSATE:
Setting this to YES enables the analyzer to adjust for any external
or internal wiring losses in the measurement of the voltage signal.
When set to YES an additional option becomes available –
WIRING LOSS.
NO
or YES
o
WIRING LOSS
Enter the estimated or measured external wiring impedance
between the point at which the analyzer’s actual voltage
measurements are being made and the point at which the desired
“virtual” voltage measurements should be made.
This is a four position, decimal point, milliohm unit of measure.
o
COMPENSATE V BURDEN
Although very small, there is a finite current flowing between the
voltage terminals of the analyzer. Depending on the wiring
method, this current may flow through the current measuring
device (internal shunt or external transducer). If it is desired to
eliminate error in current measurements caused by this excess
signal, set this option to YES, which enables the analyzer to
automatically adjust the current signal for the current flowing in
the voltage terminals.
NO
or YES
Note:
When enabled this adjusts for the actual signal flowing in
the voltage terminals, in real-time.
o
REVERSE VOLTAGE POLARITIES
Setting this to YES reverses the polarity of the voltage signal
being measured by the analyzer.
NO
or YES
o
REVERSE CURRENT POLARITIES
Setting this to YES reverses the polarity of the current signal
being measured by the analyzer.
NO
or YES
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...