154
2801/2802 Power Analyzer User Guide
Appendix C-Input & Target Specifications
Voltage Input Specifications
Input Burden
1M
Ω
|| 10pF between terminals, >10
12
Ω
|| 25pF
Voltage Lo terminal to ground
Ranges 850V
(2000Vpk)
450V
(675Vpk
135V
(200Vpk)
50V
(80Vpk)
20V
(30Vpk)
Max. Input
850Vrms, 2000Vpk max. continuous (any range),
between terminals, or either terminal to ground.
Max. External
Offsets of up to 20V may be compensated for on any
DC Offset
range, compensation is performed at the input prior to
range
selection.
Compensation
When enabled, the input may be polarity reversed
and/or adjusted to compensate for the measured
current flowing through a user entered wiring
impedance (includes the internal shunt automatically
as
applicable).
RMS Accuracy
As shown in the charts for AC, AC+DC and DC only
parameters,
filtered
(20kHz) and unfiltered.
Rectified Accuracy
1.5x RMS accuracy
Form Factor Accuracy 0.5x RMS accuracy, as a percentage of 1.0
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...