44
2801/2802 Power Analyzer User Guide
Measurements of a Power Source and Load: Method
1, Two Ind. Chnls
Source
Source
Wiring Loss
Wiring Loss
V
+
-
A
+
-
Optional
Line Switch
Load
Channel 1
(Source Connection)
Channel 2
(Load Connection)
Load
Wiring Loss
Wiring Loss
V
A
+
+
-
-
W1ohms
W2ohms
W1ohms
W2ohms
Figure 21. Measuring a Power Source and Load, Method 1, Two Ind. Chnls
Configure the channels as follows -
•
This connection method is suitable for most Independent or
Input/Output channel applications.
•
At high frequencies, the wiring length between the source and the
current and voltage measurement points should be kept as short as
possible to avoid inductive and capacitive losses causing inaccuracies.
Summary of Contents for Xitron 2801
Page 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Page 2: ...2 2801 2802 Power Analyzer User Guide...
Page 10: ...10 2801 2802 Power Analyzer User Guide...
Page 101: ...Printing Results 101 Figure 66 Configuration Printout...
Page 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Page 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Page 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Page 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Page 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Page 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Page 145: ...Sending Commands and Interrogatives 145...
Page 147: ...Appendix A Physical Specifications 147...