Front Panel Operation
27
Setting Defaults
Ø
To set defaults:
Important Note:
This action applies to all configured Results Data display lines.
Store any preferred setups before performing this step.
1. Press the
SETUP
key.
2. Press the
DISPLAY SELECT
key correlating to
Set Defaults
. The defaults will
automatically apply.
3. Press the
ENTER
key. The Results Data display will show the first four
configured lines, which are based on the preset Parameter and Setup
Measurement defaults.
For line #00 the Results Data is:
Fund
xx.xu
Vrms
A
The default
Parameter
setup is:
p = True RMSAh
1 - 1
Display as volts rms
Filter =
0.000S
For line #01 the Results Data is:
Fund
xx
.
xu
Vrms
B
The default
Parameter
setup is:
p = True RMSBh
1 - 1
Display as volts rms
Filter =
0.000S
For line #02 the Results Data is:
Greal
+ x
.
xx
B
The default
Parameter
setup is:
p = Gre
Bh01/Ah01
Display as Ratio
Filter =
0.000S
For line #03 the Results Data is:
Fund
- xxx
.
xxx
B
The default
Parameter
setup is:
p = Phs
Bh01/Ah01
Display as +/- 180
Filter =
0.000S
If you press the Nullmeter’s
SELECT
key, the default Nullmeter Parameter screen
displays. Press
ENTER
to show Nullmeter Results Data and the Nullmeter light will
reflect the reading.
The default
Nullmtr
: Parameter
setup is:
p = Phs
Bh01/Ah01
Display as +/- 180
Filter =
0.000S