78
6000-2 & -3 User Guide, Rev E
PWR—
True Power. This parameter requires an input or group of inputs and either a string
bandwidth or a pair of integers defining the harmonic range for the data. The default
units for this result is W.
QUAD—
An alternate for
IMAG
(see above).
R—
The load resistance (i.e. in-phase Volts/Current,
IMP
=
R
+ j.
X
). This parameter
requires a input and either a string bandwidth or a pair of integers defining the
harmonic range for the data. The default unit for this result is ohms.
RATIO—
An alternate for
GAIN
(see above).
RATIOI—
An alternate for
GAINI
(see above).
RATIOR—
An alternate for
GAINR
(see above).
REAL—
An alternate for
INPHS
(see above).
REA—
The load reactance (i.e. quadrature Volts/Current,
IMP
=
R
+ j.
REA
). This
parameter requires a input and either a string bandwidth or a pair of integers defining
the harmonic range for the data. The default unit for this result is ohms.
RECTIFY—
An alternate for
AVG
(see above).
RES—
An alternate for
R
(see above).
RMS—
An alternate for
AMPL
(see above).
SPAMPL—
The signal amplitude measured at the specified spectrum point for the specified
single input. The default unit for these results is
Vrms
or
Arms
as applicable.
SPFREQ—
The frequency associated with the specified spectrum point for the specified single
input. The default unit for this result is Hz.