Interface Operation
79
THDF—
The Total Harmonic Distortion of the selected signal (Volts or Current), relative to
its fundamental component. This parameter requires a single input source and a
single integer defining the highest harmonic to include in the summation. The
default unit for this result is none (i.e. a pure ratio).
THDH—
The Total Harmonic Distortion of the selected signal (Volts or Current), relative to
its total harmonic content. This parameter requires a single input source and a single
integer defining the highest harmonic to include in the summations. The default unit
for this result is none (i.e. a pure ratio).
THDT—
The Total Harmonic Distortion of the selected signal (Volts or Current), relative to
its total RMS content. This parameter requires a single input source and a single
integer defining the highest harmonic to include in the summation. The default unit
for this result is none (i.e. a pure ratio).
VAR—
Reactive or Imaginary Power. This parameter requires a input or group of inputs and
either a string bandwidth or a pair of integers defining the harmonic range for the
data. The default unit for this result is VAR.
VA—
Apparent Power. This parameter requires a input or group of inputs and either a
string bandwidth or a pair of integers defining the harmonic range for the data. The
default unit for this result is VA.
WATTS—
An alternate for
PWR
(see above).
X—
An alternate for
REA
(see above).
Z—
An alternate for
IMP
(see above).
Source and Bandwidth Strings
The
source and bandwidth
strings contain one or more of the following strings
separated by commas:
A
—
Input A.
B—
Input B.