54
6000-2 & -3 User Guide, Rev E
Imag
—
This is the quadrature component of the selected input/harmonic pair.
Phs
—
This is the phase between the selected input/harmonic pair.
Constant
—
This sets the value to what you have entered.
Reference Parameter (r)
Edit this line by pressing the corresponding
DISPLAY SELECT
key. The method of
selecting the parameter etc., is the same as for the parameter (p) line described
above.
Reference Parameter Type
This line is changed by pressing the corresponding
DISPLAY SELECT
key, using the
é
key to change the selection, then pressing the
ENTER
key. The line is only present
when the reference parameter line is present and is not set to a constant. The
available selections are:
Updated Ref. Param.
—
This selects continuous update for the reference parameter.
Fixed Ref. Param.—
This selects no continuous updating for the reference parameter, but that a previously
set value of this parameter will be used.
Fix Ref. Param. Now—
This selects that the reference parameter will be fixed at the currently measured
value. The selection changes to
Fixed Ref. Param.
after pressing the
ENTER
key.
Data Result Units
This line allows you to set the units for the result of the parameter measurement or
deviation. You may select any unit, which is compatible with the parameter or
deviation. The unit chosen is automatically set to the default when the primary
parameter is being edited. The instrument automatically provides any conversion
required.
Note that the unit is NOT automatically updated when the selected input(s) has its
mode changed between voltage and current (or
vice versa
). If you have changed the
mode, select the display line (or nullmeter select) and manually change the units,