Configuration Setup
55
specifically press the corresponding
DISPLAY SELECT
key, then use the
é
and
ê
keys to change the selection. Press the
ENTER
key when the desired unit is displayed.
The available units are:
Application
Units of Measure / Notes
Amps
“A” with optional ac, avg, dc, rms, pk, pkpk, imag or real
Volts
“V” with optional ac, avg, dc, rms, pk, pkpk, imag or real
dBV or dB
µ
V
decibel ratio to 1V or 1
µ
V respectively.
Watts
“W” with optional ac or dc
dBW or dBm
decibel ratio to 1W or 1mW respectively
VA
with optional dc, ac or pk
VAR
with optional ac
Horse power
“HP”
%
percentage
dB
decibel
±
180° or 0-360°
Half or full circle
Grad
±
200 or 0-400
Radians
±π
or 0-2
π
π
radians
±
1 or 0-2
Hertz
“Hz”
Cycles per second
“cps”
Revolutions per min.
“rpm”
Seconds
“sec”, inverse Hz
Hours
“Hr”
Resistance
“
Ω
”, with optional dc, ac, imag or real
Conductance
“S”, with optional dc, ac, imag or real
Comparison Limit
If you have selected a limit comparison to be performed, then this line becomes
visible. The constant value set in this line sets the amount by which the primary
parameter may deviate from the reference parameter before the displayed status
changes from
PASS
to either
HIGH
(p > r + limit) or
LOW
(p < r: limit). The limit
may be changed by pressing the corresponding
DISPLAY SELECT
key, using the
numeric keys to enter the desired limit, pressing the
ENTER
key when the desired
limit has been entered.